PA-II-SD 68c63acc07a9 | 2020-10-22 | Bouldin Corn Air Quality Sensor | Removed | Lab none | Removed from BC and brought back to the lab. The microSD card is hard to seat properly into its slot. It was replaced by sn 84f3ebd5ffb1. |
PAR-LITE 990031 | 2020-10-26 | Lab none | Repair | Lab none | The cable insulation on this sensor was rotted. I cut the cable off near the sensor and spliced on a new cable. I also added an aluminum bracket to provide strain relief and used liquid electrical tape to seal the hole where the cable enters the sensor. |
PAR-LITE 990025 | 2020-10-26 | Lab none | Repair | Lab none | The cable insulation on this sensor was rotted. I cut the cable off near the sensor and spliced on a new cable. I also added an aluminum bracket to provide strain relief and used liquid electrical tape to seal the hole where the cable enters the sensor. The white wire (+) had a nick in its insulation right at the point where it entered the sensor that I covered with heat shrink. |
SRS-Nr NDVI 943203017 | 2020-10-22 | Tonzi | Removed | Lab none | Removed from the Tonzi floor met because it was no longer being used - tree NDVI pointing to the sky doesn't work |
eosFD FD20180048 | 2020-10-22 | Lab none | Installed | Tonzi-SoilCO2 Forced Diffusion Tree 1 | Installed at the under tree location at Tonzi to replace sn FD20180068 which has been reading zero CO2 concentration. |
eosFD FD20180068 | 2020-10-22 | Tonzi-SoilCO2 Forced Diffusion Tree 1 | Removed | Lab none | Removed from the under tree location at Tonzi because its co2 sensor has been reading zero. Replaced by sn FD20180048 |
Line Quantum LQ1570 | 2001-05-07 | Vaira | Moved | Tonzi none | Moved from Vaira to Tonzi |
BB-Fab BB-TF | 2003-12-19 | Tonzi-Floor Battery Bank | Upgrade | Tonzi-Floor Battery Bank | Removed four batteries from the floor system and install eoght new Excide NC27 95Ahr batteries. Treated all terminals with antiseize compound. The four batteries removed will be used at Vaira. |
BB-Fab BB-VR | 2003-12-19 | Vaira Battery Bank Main | Upgrade | Vaira Battery Bank Main | Cleaned battery terminals and added four batteries from the Floor site for a total of 12 at Vaira. |
LI-7500A 0035 | 2003-10-02 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0073 for calibration. |
LI-7500A 0073 | 2003-10-02 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0035. |
LI-7500A 0418 | 2003-10-02 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0041 for calibration. |
LI-7500A 0041 | 2003-10-02 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0418. |
LI-7500A 0042 | 2003-10-02 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0065 for calibration. |
LI-7500A 0065 | 2003-10-02 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0042. |
RPI0W RPI0W-000 | 2020-10-20 | Lab none | Tested | Lab none | Tested in the lab using spare parts from sn RPI0W-004. The board overall seems fine, but the board cannot connect to the camera, so the board is not going to be useful to run cameras in the future. The camera and camera cable seem good, so we have a spare camera and camera cable here. Running updates did not seem to change anything. I also changed the clock +7 min. |
RPI0W RPI0W-000 | 2020-10-05 | Sherman Wetland-Temp1 Camera | Removed | Lab none | Removed from SW Temp Tower and brought back to the lab because the camera was not connecting. |
RPI0W RPI0W-000 | 2020-01-30 | East Pond Camera | Moved | Sherman Wetland-Temp1 Camera | Removed with the decommissioning of the EP tower and installed at the new SW Temporary site 1250m upwind of the Permanent Tower. |
LI-7500A 0042 | 2003-06-20 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0065 for calibration. |
LI-7500A 0065 | 2003-06-20 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0042. |
LI-7500A 0035 | 2003-06-20 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0041 for calibration. |
LI-7500A 0041 | 2003-06-20 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0035. |
LI-7500A 0418 | 2003-06-20 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0073 for calibration. |
LI-7500A 0073 | 2003-06-20 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0418. |
WMP 1352 000024 | 2003-05-30 | Lab none | Installed | Tonzi-Floor Anemometer Floor | Installed on Tonzi-Floor to replace sn 000019 that was having problems - not sending data. |
WMP 1352 000019 | 2003-05-30 | Tonzi-Floor Anemometer Floor | Removed | Lab none | Removed from Tonzi-Floor due to no data. On power up it would send some data and then quit. |
LI-7500A 0073 | 2003-05-30 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira due to bad power connector on the box. |
LI-7500A 0065 | 2003-05-30 | Tonzi-Floor Gas Analyzer CO2/H2O | Moved | Vaira Gas Analyzer CO2/H2O | Moved from Tonzi-Floor to Vaira because the Floor sonic was dead and the Vaira 7500 box had a bad power connector |
LI-7500A 0073 | 2003-05-09 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0042 for calibration. |
LI-7500A 0042 | 2003-05-09 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0073. |
LI-7500A 0418 | 2003-05-09 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0065 for calibration. |
LI-7500A 0065 | 2003-05-09 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0418. |
LI-7500A 0035 | 2003-05-09 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0041 for calibration. |
LI-7500A 0041 | 2003-05-09 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0035. |
LI-7500A 0042 | 2003-04-06 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0418 for calibration. |
LI-7500A 0418 | 2003-04-06 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0042. |
LI-7500A 0065 | 2003-03-17 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0073 for calibration. |
LI-7500A 0073 | 2003-03-17 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0065. |
LI-7500A 0041 | 2003-03-17 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0035 for calibration. |
LI-7500A 0035 | 2003-03-17 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0041. |
BB-Fab BB-TT | 2003-02-13 | Tonzi-Tower Battery Bank | Upgrade | Tonzi-Tower Battery Bank | Four existing batteries removed for use at Vaira. 12 new batteries from Ione Autoparts installed |
BB-Fab BB-VR | 2003-02-13 | Vaira Battery Bank Main | Upgrade | Vaira Battery Bank Main | Added four batteries removed from Tonzi-Tower. Now a total of eight batteries. |
HMP45AC W0510026 | 2003-01-21 | Lab none | Installed | Tonzi none | Installed at Tonzi to replace sn U3030032 for calibration. |
HMP45AC U3030032 | 2003-01-21 | Tonzi none | Removed | Lab none | Removed from Tonzi. Replaced by sn W0510026 |
WMP 1352 000019 | 2003-01-07 | Lab none | Installed | Tonzi-Floor Anemometer Floor | Installed at Tonzi-Floor to replace sn 000021. |
WMP 1352 000021 | 2003-01-07 | Tonzi-Floor Anemometer Floor | Removed | Lab none | Removed from Tonzi-Floor. Replaced by sn 000019. |
BB-Fab BB-TF | 2001-01-03 | Lab none | Installed | Tonzi-Floor Battery Bank | Initial installation with four solar panels and three deep cycle batteries. |
LI-7500A 0042 | 2000-10-12 | Lab none | Calibration | Lab none | Zero with N2 was 2.0, after running several weeks in the lab.Cal gas 402.25 read 411.1 ppm at P=101.31 kPa and T 23.17 C. Licor Dewpoint sensor read 14.57 Tdew. Vaisala standard read 13.9 C. We are using the Vaisala as the standard for the experiment. Rezeroed and spanned D/A #1, H2O mmol m-3 0v=0 5 v=1000 D/A #2, CO2 mmol m-3 0v=13 mmol m-3 5v=20 mmol m-3 |
ML2x 54/020 | 2000-10-12 | Lab none | Calibration | Lab none | Water: 1.1367V vs air (zero): -0.00101V |
ML2x 54/019 | 2000-10-12 | Lab none | Calibration | Lab none | Water: 1.1317V vs air (zero): -0.000034V |
ML2x 54/017 | 2000-10-12 | Lab none | Calibration | Lab none | Water: 1.1294V vs air (zero): -0.00101V |
ML2x 54/016 | 2000-10-12 | Lab none | Calibration | Lab none | Water: 1.1403V vs air (zero): -0.00067V |
ML2x 54/015 | 2000-10-12 | Lab none | Calibration | Lab none | Water: 1.1381V vs air (zero): 0.00318V |
LI-7500A 0041 | 2002-06-06 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0035. |
LI-7500A 0042 | 2002-12-23 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0035 for calibration. |
LI-7500A 0035 | 2002-12-23 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0042. |
LI-7500A 0041 | 2002-12-23 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0073 for calibration. |
LI-7500A 0073 | 2002-12-23 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0041. |
LI-7500A 0065 | 2002-12-23 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0418 for calibration. |
LI-7500A 0418 | 2002-12-23 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0065. |
LI-7500A 0042 | 2002-09-20 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0418 for calibration. |
LI-7500A 0418 | 2002-09-20 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0042. |
LI-7500A 0065 | 2002-09-20 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0073 for calibration. |
LI-7500A 0073 | 2002-09-20 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0065. |
LI-7500A 0041 | 2002-09-20 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0035 for calibration. |
LI-7500A 0035 | 2002-09-20 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0041. |
LI-7500A 0418 | 2002-08-15 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0042 for calibration. |
LI-7500A 0042 | 2002-08-15 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0418. |
LI-7500A 0035 | 2002-08-15 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0041 for calibration. |
LI-7500A 0041 | 2002-08-15 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0035. |
LI-7500A 0073 | 2002-08-15 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0065 for calibration. |
LI-7500A 0065 | 2002-08-15 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0073. |
eosFD FD20190028 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection and repair |
eosFD FD20190028 | 2020-09-28 | Lab none | Tested | Lab none | No pump sounds (orange light inside comes on) and both CO2 channels read the same. The flux is zero. Probably dead pump. I'm tempted to take this one apart and look for a loose wire but it's less than a year old and I don't want to void any warranty. |
eosFD FD20160001 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for evaluation and repair. |
eosFD FD20160001 | 2020-09-28 | Lab none | Tested | Lab none | The pump is running but both CO2 signals are just noise. Probably another dead IRGA. Also there's a scratch in oneof the atmospheric diffusion membranes. |
GP-001 GP20150101 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection |
GP-001 GP20140016 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection |
GP-001 GP20160069 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection |
GP-001 GP20160096 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection |
GP-001 GP20140017 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection |
GP-001 GP20140019 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection |
GP-001 GP20150109 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection |
GP-001 GP20140018 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection |
GP-001 GP20140088 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection. No record of this sensor ever being used in the field but it was in the pile of bad sensors. |
GP-001 GP20160096 | 2020-10-05 | East End CO2 probe | Removed | Lab none | Removed from EE because it still rails high in air and in water. Will send it back to the factory |
BB-Fab BB-EE | 2020-10-05 | East End Battery Bank | Replaced | East End Battery Bank | Replaced old battery bank with 4 brand-new sealed AGM batteries. The terminals have been sprayed with anti-corrosion grease. |
GP-001 GP20160096 | 2019-10-02 | Lab none | Installed | East End CO2 probe | Installed at EE off the north side of the tower. |
LI-7500A 0042 | 2002-07-19 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0073 for calibration. |
LI-7500A 0073 | 2002-07-19 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0042. |
LI-7500A 0041 | 2002-07-11 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0035 for calibration. |
LI-7500A 0035 | 2002-07-11 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 0041. |
LI-7500A 0065 | 2002-07-11 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0042 for calibration. |
LI-7500A 0042 | 2002-07-11 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0065. |
LI-7500A 0073 | 2002-06-20 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0065 for calibration. |
LI-7500A 0065 | 2002-06-20 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0073. |
LI-7500A 0073 | 2002-05-09 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0035 for calibration. |
LI-7500A 0035 | 2002-05-09 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 0073. |
LI-7500A 0065 | 2002-05-09 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0042 for calibration. |
LI-7500A 0042 | 2002-05-09 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0065. |